CVE-2020-12040
CVSS score9.8 CRITICAL
EPSS probability0.95%
CISA KEVNot currently listed
Published2020-06-29
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Weakness classification
- ics-cert@hq.dhs.gov: CWE-319
- nvd@nist.gov: CWE-319
Primary references
- ics-cert@hq.dhs.gov — Third Party Advisory, US Government Resource
- af854a3a-2127-422b-91ae-364da2661108 — Third Party Advisory, US Government Resource