CVE-2020-12037

CVSS score7.5 HIGH
EPSS probability0.48%
CISA KEVNot currently listed
Published2020-06-29
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Description

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Weakness classification

  • ics-cert@hq.dhs.gov: CWE-259
  • nvd@nist.gov: CWE-319