VULNERABILITY INTELLIGENCE

CVE-2019-6538

CVSS score9.3 CRITICAL
EPSS probability0.90%
CISA KEVNot currently listed
Published2019-03-25
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Weakness classification

  • CWE-284: Improper Access Control
  • CWE-306: Missing Authentication for Critical Function
  • CWE-862: Missing Authorization

Primary references