VULNERABILITY INTELLIGENCE
CVE-2019-6538
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Weakness classification
- CWE-284: Improper Access Control
- CWE-306: Missing Authentication for Critical Function
- CWE-862: Missing Authorization
Primary references
- ics-cert@hq.dhs.gov — Third Party Advisory, VDB Entry
- ics-cert@hq.dhs.gov — Mitigation, Third Party Advisory, US Government Resource
- af854a3a-2127-422b-91ae-364da2661108 — Third Party Advisory, VDB Entry
- af854a3a-2127-422b-91ae-364da2661108 — Mitigation, Third Party Advisory, US Government Resource