VULNERABILITY INTELLIGENCE
CVE-2018-8868
CVSS score6.2 MEDIUM
EPSS probability0.33%
CISA KEVNot currently listed
Published2018-07-03
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
Description
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Weakness classification
- CWE-749: Exposed Dangerous Method or Function
Primary references
- ics-cert@hq.dhs.gov
- ics-cert@hq.dhs.gov — Third Party Advisory, US Government Resource
- af854a3a-2127-422b-91ae-364da2661108 — Third Party Advisory, US Government Resource