VULNERABILITY INTELLIGENCE

CVE-2018-5438

CVSS score6.3 MEDIUM
EPSS probability0.44%
CISA KEVNot currently listed
Published2018-03-20
Last modified2026-06-17
CVSS vectorCVSS:3.0/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:N

Description

Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists when using ISCV together with an Electronic Medical Record (EMR) system, where ISCV is in KIOSK mode for multiple users and using Windows authentication. This may allow an attacker to gain unauthorized access to patient health information and potentially modify this information.

Weakness classification

  • nvd@nist.gov: CWE-613

Primary references