VULNERABILITY INTELLIGENCE

CVE-2018-14781

CVSS score5.3 MEDIUM
EPSS probability0.71%
CISA KEVNot currently listed
Published2018-08-13
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N

Description

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Weakness classification

  • CWE-287: Improper Authentication
  • CWE-294: Authentication Bypass by Capture-replay