VULNERABILITY INTELLIGENCE
CVE-2017-5149
CVSS score8.9 HIGH
EPSS probability1.24%
CISA KEVNot currently listed
Published2017-02-13
Last modified2026-06-17
CVSS vectorCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
Description
An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
Weakness classification
- CWE-476: NULL Pointer Dereference
Primary references
- ics-cert@hq.dhs.gov — Third Party Advisory, VDB Entry
- ics-cert@hq.dhs.gov — Mitigation, Third Party Advisory, US Government Resource
- af854a3a-2127-422b-91ae-364da2661108 — Third Party Advisory, VDB Entry
- af854a3a-2127-422b-91ae-364da2661108 — Mitigation, Third Party Advisory, US Government Resource