VULNERABILITY INTELLIGENCE

CVE-2017-14018

CVSS score4.8 MEDIUM
EPSS probability0.36%
CISA KEVNot currently listed
Published2017-12-05
Last modified2026-06-17
CVSS vectorCVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L

Description

An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.

Weakness classification

  • CWE-287: Improper Authentication