VULNERABILITY INTELLIGENCE

CVE-2017-12725

CVSS score5.6 MEDIUM
EPSS probability0.83%
CISA KEVNot currently listed
Published2018-02-15
Last modified2026-06-17
CVSS vectorCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Weakness classification

  • CWE-798: Use of Hard-coded Credentials

Primary references