VULNERABILITY INTELLIGENCE
CVE-2015-3952
CVSS score7.5 HIGH
EPSS probability1.01%
CISA KEVNot currently listed
Published2019-03-25
Last modified2026-06-17
CVSS vectorCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Description
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Weakness classification
- CWE-200: Exposure of Sensitive Information to an Unauthorized Actor
- CWE-312: Cleartext Storage of Sensitive Information
Primary references
- ics-cert@hq.dhs.gov — Mitigation, Third Party Advisory, US Government Resource
- af854a3a-2127-422b-91ae-364da2661108 — Mitigation, Third Party Advisory, US Government Resource